background image

B-TEK SCALES, LLC    1510 METRIC AVE. SW CANTON, OH 44706    TOLL FREE: 800.266.8900    FAX: 330.471.8909    WWW.B-TEK.COM

Introduction

This technical guide will provide instructions on how to install and set up the analog output in a T403S, T405S, 

and T419S indicator. The Analog Option is P/N: 825-250028. The T-Series analog output uses a 16-bit D/A and 

provides approximately 16K divisions of accuracy over the temperature range of the product (-10 to 40C)

Installation

Install the analog output option PCB following the instructions supplied with the option kit or refer to the  

appropriate Service manual. Set the jumpers P2 thru P8 for the desired current or voltage output using the 

chart below. Connect the appropriate terminals on TB5 for voltage or current output to the PLC or other analog 

input device. TB5 - 2 is Ground for Voltage outputs or Current Return for Current outputs. Refer to the Min/Max 

Load Resistance ratings noted in the Analog Output Chart. Also check to make sure all S1 switches are in the 

OFF position.

Install and Configure a T-Series Analog Option Card

Отзывы: