Analog Output Protocol Setup
To complete the configuration requires setup of a protocol in the
PORTS >
PROTCL
menu to output the
desired scale information to the Analog Option port. Use any available protocol
PROT 1
thru
5
1. From the previous step in Calibration procedure press
UNITS
until
PROT
is displayed
2. Perform the required steps to setup a protocol with the following settings
a.
TYPE x = B-CAST
provides continuous output
b.
BIND x = OPTION
protocol settings apply to the Analog option card
c.
ATTR x > PRNFT = 32
a pre-defined print format for the analog option
d.
ATTR x > INHIB = OFF
continuous output not disabled by motion settings
e.
ATTR x > RATE = 20
output data at the fastest rate setting
3. After completing the above protocol setup press
TARE
until
SAVE NO
is displayed
4. Press
UNITS >
SAVEYES
is displayed
5. Press
ZERO >
BUSY
is displayed. The indicator will reboot and return to normal weighing
Install and Configure a T-Series Analog Option Card