Analog Output Protocol Setup

To complete the configuration requires setup of a protocol in the 

PORTS > 

PROTCL

 menu to output the  

desired scale information to the Analog Option port. Use any available protocol 

PROT 1

 thru 

5

1.  From the previous step in Calibration procedure press 

UNITS

 until 

PROT

 is displayed

2.  Perform the required steps to setup a protocol with the following settings

 a. 

TYPE x = B-CAST

  

provides continuous output

 b. 

BIND x = OPTION

  

protocol settings apply to the Analog option card

 c. 

ATTR x > PRNFT = 32

 

a pre-defined print format for the analog option

 d. 

ATTR x > INHIB = OFF

  

continuous output not disabled by motion settings

 e. 

ATTR x > RATE = 20

  

output data at the fastest rate setting

3.  After completing the above protocol setup press 

TARE

 until 

SAVE NO

 is displayed

4. Press 

UNITS > 

SAVEYES

 is displayed

5. Press 

ZERO > 

BUSY

 is displayed. The indicator will reboot and return to normal weighing

Install and Configure a T-Series Analog Option Card