Purpose
Read or set the current pressure stability limit.
Command
Query
“SS
limit
”
“SS?”
Classic “SS=
limit
”
“SS”
Optional Suffix
“n”
A suffix may be used instead of “
RptLabel
”
‘1’ Specify the Hi Q-RPT
‘2’ Specify the Lo Q-RPT
Arguments
Limit:
The stability limit in the current pressure unit of measure.
Remarks
The stability limit can be read and set as a pressure. The stability limit is used as the
Ready/Not Ready
criteria..
Example
(enhanced)
Cmd sent:
Query reply:
“SS .1”
“0.10 psi/s”
Example
(enhanced)
Cmd sent:
Query reply:
“SS? .1”
“0.10 psi/s”
Example
(classic)
Sent:
Reply:
“SS=.1”
“0.10 psi/s”
Errors ERR#
6:
ERR# 10:
The argument was invalid.
Invalid or missing command suffix.
See Also
“SS%”
3.8.3, 3.1.2
TIME
Purpose
Read or set the RPM4-E-DWT internal clock.
Command
Query
“TIME
hh:mmXX
”
“TIME?”
Classic “TIME=
hh:mmXX
”
“TIME”
Arguments
hh:mm:
XX:
The time in a 12 hour format using a colon delimiter
”am” or “pm”
Example
(enhanced)
Cmd sent:
Query reply:
“TIME 12:52PM”
“12:52pm”
(No reply if GPIB-488)
Example
(enhanced)
Cmd sent:
Query reply:
“TIME? 12:52PM”
“12:52pm”
Example
(classic)
Sent:
Reply:
“TIME=12:52PM”
“12:52pm”
Errors
ERR# 7:
Missing or improper program message argument(s).
See Also
“DATE”
3.9.5.3
UCOEFn
Purpose
To identify the coefficient used to convert 1 Pascal to the current pressure unit of measure.
Query “UCOEF
n
?”
Classic “UCOEF
n
”
Optional Suffix
“n”
A suffix may be used instead of “
RptLabel
”
‘1’ Specify the Hi Q-RPT
‘2’ Specify the Lo Q-RPT
Remarks
The RPM4-E-DWT handles all pressure values internally in Pascal. The coefficient replied is
equivalent of 1 Pa in the current pressure unit of measure. This program message allows the
user to convert pressures
Example
(enhanced)
Query sent:
Query reply:
“UCOEF?”
“0.0010000000”
Example
(classic)
Sent:
Reply:
“UCOEF”
“0.0010000000”
See Also
3.7.3, 3.9.6, 7.1.1
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...