E-DWT-10000-AF OPERATION AND MAINTENANCE MANUAL
© 2008 DH Instruments, a Fluke Company
Page 34
RPT
Allows pressure measurement from the RPM4-E-DWT’s two Q-RPTs to be
displayed simultaneously (see Section 3.7.7.4).
HI/LO
Records and displays maximum and minimum pressures measured (see
Section 3.7.7.5). This function is used to keep track of the minimum and
maximum pressure observed in a system over a period of time or to monitor if
a pressure min/max limit has been exceeded.
FREEZE
Captures and displays the pressure measured by the active range of
RPM4-E-DWT when the
[ENT]
key is pressed (see Section 3.7.7.6). This
function is useful to record the pressure present at the time of an operator
observed trigger event, for example when the needle of an analog gauge is on
the nominal point or when a switch activates.
CLEAN
Blanks out the second line of the display (see Section 3.7.7.7). This function is
used when a simple display of pressure measured by the RPM4-E-DWT active
range with minimal additional information is desired.
OPERATION
To set the DISPLAY function press
[DISPLAY]
from the main run screen.
The display is:
1Avg 2Rate 3Dev 4RPT
5Hi/Lo 6Freez 7Cln
The cursor is on the active DISPLAY function. Selecting a DISPLAY function returns to the
main run screen with the selected function active. See Sections 3.7.7.1 to 3.7.7.7 for
information on each of the DISPLAY choices.
The DISPLAY selection is NOT range specific. A DISPLAY selection made in one range
applies to all ranges.
The default DISPLAY function is Rate which causes the second line of the display to show “R”
followed by the current rate of change of pressure in current pressure units per second.
3.7.7.1
AVG (AVERAGE)
PURPOSE
To activate the Average DISPLAY and/or adjust the period of time over which
averaging occurs.
See
Section
3.7.7.
Principle.
OPERATION
To access the Average DISPLAY, press [DISPLAY], <1Avg>. The display is:
1. Edit field for averaging period in seconds.
Default is 20. Minimum 1, maximum 999.
Averaging Period:
20 s
Edit the averaging time period if desired. Press
[ENT]
to return to the main run
screen with the Average DISPLAY active.
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...