E-DWT-10000-AF OPERATION AND MAINTENANCE MANUAL
© 2008 DH Instruments, a Fluke Company
Page 48
Use the cursor control keys to set the desired level of resolution. Press
[ENT]
to return to the
MAIN RUN screen with the new resolution setting active or
[ESC]
to make no changes.
The resolution setting is range specific. A resolution setting made in one range does NOT
affect other ranges.
The maximum resolution setting is limited in AutoRange ranges that are less than 10% of
the Q-RPT default full scale.
3.8.3
<3STAB> (STABILITY)
PURPOSE
To view and/or adjust the stability test that is the Ready/Not Ready criterion for the active
Q-RPT and range.
See
Section
3.1.2 Pressure Ready/Not Ready.
PRINCIPLE
The RPM4-E-DWT continuously monitors the rate of change of pressure measured by the
active Q-RPT and compares this rate to the active stability limit to make a
Ready/Not Ready
determination (see Section 3.1.2). The stability function allows the stability limit to be
adjusted by the user to increase or decrease the stability required for a
Ready
condition to
occur.
Default stability when using AutoRange is + 0.01% FS of active range span. Default
stability when AutoRanging as part of an AutoTest is 10% of the DUT tolerance (see Section 3.7.5,
Table 4).
The stability limit is separate and different from the Rate DISPLAY function (see Section
3.7.7.2) which allows the current rate of change of pressure to be displayed.
OPERATION
To access the stability setting adjustment, press
[SETUP]
,
<3stab>
. The display is:
Stability limit: Hi
1.0 psi/s
Edit the desired stability limit setting.
[ENT]
activates the stability limit for the active range
and returns to the main run screen Press
[ESC]
to return to the MAIN RUN screen with no
change to the stability limit.
The stability setting is range specific. A stability setting made in one range does NOT
affect other ranges.
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...