![DH Instruments E-DWT-10000-AF Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/dh-instruments/e-dwt-10000-af/e-dwt-10000-af_operation-and-maintenance-manual_2492026045.webp)
3. OPERATION
Page 35
© 2008 DH Instruments, a Fluke Company
With the Average DISPLAY active the main run screen is:
1. Average measured over last completed
averaging period.
2. Standard deviation of last completed
averaging period.
3.
Countdown in seconds until completion of on-
going averaging period.
2007 psi g Hi
δ
0.2 18Avg
The Average DISPLAY has a second screen that allows the instantaneous
pressure readings to be viewed while an averaging cycle is running.
The instantaneous Average screen is:
1. Instantaneous pressure values at RPM4-E-
DWT’s normal update rate.
2.
Current rate of change of pressure in pressure
units/second.
3.
Countdown in seconds until completion of on-
going averaging period.
2006 psi g Hi
R 0.1/s 18Avg
The
[+/-]
key toggles between the main run Average screen and the
instantaneous values Average screen.
Pressing [ENT] while in the Average DISPLAY aborts the current averaging
period and causes a new one to begin. [ENT] can thus be use to trigger a new
averaging period on demand.
In the Average DISPLAY the Ready/Not Ready indication applies to the result
of the previous averaging period (see Section 3.1.2). Ready indicates that all
readings during the previous averaging period met the stability criterion. Not Ready
indicates that one or more readings were outside of the stability criterion.
Changing the pressure unit of measure, measurement mode or range while the
averaging screen is active, starts a new averaging period.
To go to a DISPLAY other than Average, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.7.7).
3.7.7.2
Rate
PURPOSE
To activate the Rate DISPLAY.
See
Section
3.7.7,
Principle.
The Rate run screen accesses the AutoTest function (see Section 3.7.11) to
AutoRange RPM4-E-DWT based on DUT characteristics. To access AutoTest, press
[ENT] from the Rate run screen.
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...