![DH Instruments E-DWT-10000-AF Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/dh-instruments/e-dwt-10000-af/e-dwt-10000-af_operation-and-maintenance-manual_2492026046.webp)
E-DWT-10000-AF OPERATION AND MAINTENANCE MANUAL
© 2008 DH Instruments, a Fluke Company
Page 36
OPERATION
To activate the Rate DISPLAY press
[DISPLAY]
,
<2Rate>.
Pressing
<2Rate>
returns to the main run screen with the Rate DISPLAY active.
With the Rate DISPLAY active the main run screen is:
1.
Current rate of change of pressure in current
pressure unit per second.
8455 psi g Hi
R 0.1/sec
The Rate DISPLAY is different and separate from the stability setting which is
used to set the stability criterion on which the
Ready/Not Ready
indication is based
(see Sections 3.8.3 and 3.1.2). The Rate DISPLAY only causes the current rate of
change to be displayed and has NO affect on the stability setting or the Ready/Not
Ready condition.
The Rate DISPLAY is the default RPM4-E-DWT main run screen display.
To go to a DISPLAY other than Rate, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.7.7).
3.7.7.3
Dev (Deviation)
PURPOSE
To activate the Deviation DISPLAY and/or set the deviation target value.
See
Section
3.7.7,
Principle.
OPERATION
To activate the Deviation DISPLAY press
[DISPLAY]
,
<3Dev>
. The display is:
Target:
9000 psi g
Edit the desired target value. Pressing
[ENT]
returns to the main run screen with
the Deviation DISPLAY active using the entered target value.
The target value is the value from which deviations (D) are measured by the
Deviation DISPLAY following:
D = current pressure - target pressure
With the Deviation DISPLAY active the main run screen is:
1.
Deviation of current pressure from the target
value.
2. Target
value.
8998 psi a zH3
D -2 T9000.0
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...