File AutoTest parameters are stored in definition files.
[SETUP]
,
<5AutoTest>
,
<2file>
allows File AutoTest definition files to be viewed and edited. Up to 40
File AutoTest definition files with up to 21 test points in each file can be created.
A File AutoTest definition file defines:
•
The identifying number of the File AutoTest definition file.
•
Pressure unit of measure.
•
DUT full scale in pressure unit of measure and measurement mode.
•
DUT tolerance (% FS).
•
Number of test points or pressure targets to include in the AutoTest.
•
Numerical value of each test point or pressure target.
See Section 3.7.11,
OPERATION
for information on running a Quick or File
AutoTest.
See Section 3.8.5.1 for information on viewing the data files that result from
running a Sequence.
OPERATION
See
Section
3.7.11, OPERATION, for information on RUNNING a Quick or File
AutoTest.
To access the AutoTest definition viewing and editing functions press
[SETUP]
,
<5ATest>
,
<2file>
. Select
<1view>
to view an AutoTest definition file. Select
<2edit>
to edit an AutoTest definition file.
When
<2edit>
is selected the display is:
Edit File AutoTest #
1
Enter the number (1 - 40) of the File AutoTest to be edited.
To create a File AutoTest definition file similar to an existing definition file
without entering all the definition information, edit an existing definition file and at
the end of the editing process, save it as a different file number.
The display is:
1psi 2MPa 3kPa
4bar
The unit of measure selection screen is the current
[UNIT]
selection screen (see
Section 3.7.3). Select the desired pressure measurement unit for the AutoTest.
The display is:
DUT full scale:
6000 psi g File1
Enter the full scale pressure of the DUT to be calibrated with this File AutoTest.
The value entered will be used along with the DUT tolerance to AutoRange
RPM4-E-DWT settings when the File AutoTest is run (see Section 3.7.5,
PRINCIPLE
).
The display is:
DUT tolerance:
0.25% FS File1
Enter the tolerance in % of full scale (measurement uncertainty, precision or
accuracy as desired) of the DUT to be calibrated using this File AutoTest. The
value entered will be used along with the DUT full scale to determine
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...