![DH Instruments E-DWT-10000-AF Скачать руководство пользователя страница 96](http://html1.mh-extra.com/html/dh-instruments/e-dwt-10000-af/e-dwt-10000-af_operation-and-maintenance-manual_2492026096.webp)
E-DWT-10000-AF OPERATION AND MAINTENANCE MANUAL
© 2008 DH Instruments, a Fluke Company
Page 86
RESET
Purpose
Reset the user’s settings to factory defaults.
Command “RESET”
Classic “RESET”
Remarks
The RPM4-E-DWT has user settings (units, resolution, etc.) that can be reset to factory
defaults. The remote “RESET” program message corresponds to the front panel “Reset - Sets”.
System calibration coefficients and communications settings are not affected. The reset cycle
takes up to 3 seconds to complete. Remote communications should not take place during
this period.
Example
(enhanced)
Cmd sent:
Reply:
“RESET”
“RESET”
(no reply if IEEE-488)
Example
(enhanced)
Cmd sent:
Reply:
“RESET?”
“RESET”
Example
(classic)
Sent:
Reply:
“RESET”
“RESET”
See Also
3.9.9.1
RPTn
Purpose
Read the available Q-RPT identification data.
Query
“RPT
n
”
Classic
“RPT
n
”
Optional Suffix
“n”
A suffix may be used instead of “
RptLabel
”
‘1’ Specify the Hi Q-RPT
‘2’ Specify the Lo Q-RPT
Remarks
Two internal Q-RPTs are available for use in an RPM4-E-DWT. You can query the
RPM4-E-DWT for information about each particular Q-RPT. The Q-RPT data is returned in the
following format:
RPTLabel, Q-
RPTLocator, Serial#, RngGa, RngAbs, Q-RPTMode
RngLabel:
RPT type label. This label identifies the Q-RPT type and range. This is
the same label used on the RPM4-E-DWT front panel screen.
RptLocator:
Text field identifying the Q-RPT position in the RPM4-E-DWT system.
“IH” identifies this as an internal, Hi Q-RPT
“IL” identifies this as an internal, Lo Q-RPT
Serial#:
The serial number of the Q-RPT.
RngGa:
The Q-RPT default gauge mode range in the current pressure unit.
RngAbs:
The Q-RPT default absolute mode range in the current pressure unit.
“NONE” appears in the field if the Q-RPT is a Gxxx, BGxxx or Axxx with
absolute and negative gauge modes OFF.
RptMode:
‘G’ if Q-RPT is gauge Gxxx or Axxx with absolute and negative gauge
modes OFF and supports only gauge measurement mode.
Example
(enhanced)
Cmd sent:
Query reply:
“RPT2?”
(Get information on the Lo Q-RPT)
“A7M, IL, 82345, 1000, NONE,G”
Example
(classic)
Cmd sent:
Query reply:
“RP1”
(Get information on the HL Q-RPT)
“A70M, HL, 82349, 10000, NONE,G”
Errors ERR#4:
ERR# 10:
External device not detected.
Invalid suffix.
See Also
3.1.1
SCRSAV
Purpose
Read or set the front panel display screen saver activation time.
Command
Query
“SCRSAV
n
”
“SCRSAV?”
Classic “SCRSAV=
n
”
“SCRSAV”
Arguments
n:
The inactivity period (0 – 99 minutes) after which screen saver activates.
Default “SCRSAV
10”
Remarks
The RPM4-E-DWT front panel will dim after a period of keyboard and remote inactivity.
Setting this value to ‘0’ disables this feature.
Example
(enhanced)
Cmd sent:
Query reply:
“SCRSAV 30”
“30”
(No reply if GPIB-488)
Example
(enhanced)
Cmd sent:
Query reply:
“SCRSAV? 30”
“30”
Example
(classic)
Sent:
Reply:
“SCRSAV=30”
“30”
Errors
ERR# 6
The argument was invalid.
See Also
3.9.5.1
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...