E-DWT-10000-AF OPERATION AND MAINTENANCE MANUAL
© 2008 DH Instruments, a Fluke Company
Page 44
•
<DUT full scale:>:
Enter the full scale pressure (maximum pressure) of the
DUT in the measurement mode and pressure units that have been specified
in the previous two steps.
•
<DUT tolerance:>:
Enter the DUT measurement uncertainty, accuracy or
precision specification as desired in
±
% FS (percent of full scale). This
value will be used to determine appropriate RPM4-E-DWT resolution and
stability test settings.
•
<Test increment:>:
Enter the test increment in % FS of the DUT.
When executing the test, the test increments will be from zero to the DUT full
scale, dividing by the value of the increment. If a non-even increment value is
entered, the actual increments will be rounded to the lower number of
increments. For example, if the DUT full scale is 10 000 psi and the increment
specified is 42 %, the Sequence will run points of 0, 5 000 and 10 000 psi
because 42 % goes into 100 less than three but more than two times.
•
<Test point sequence:>:
Select <1up> for the AutoTest sequence to run
from zero to the DUT full scale; <2down> for the sequence to start at the
DUT full scale and run down to zero, <3u&d> for the sequence to run from
zero to the DUT full scale and back to zero; <4d&u> for the sequence to run
from the DUT full scale down to zero and back to the DUT full scale.
After the last test definition entry, a screen is offered for the user to check that
the correct test has been defined or selected and that it does not exceed the
range of the device or system being tested:
1. Indication of minimum and maximum
pressure included in the sequence which is
about to run.
ENTER to run psi g
0 to 5000
Pressing
[ESC]
returns to the AutoTest selection screen without AutoRanging the
RPM4-E-DWT and without sequence execution. Pressing
[ENT]
causes the
RPM4-E-DWT to AutoRange based upon the DUT characteristics (see Section 3.7.5)
and the AutoTest proceeds to text execution (see Section 3.7.11.2). To AutoRange the
RPM4-E-DWT based on DUT characteristics but not run a test sequence,
[ESC]
right
after AutoRanging completes (at the
<Set valve for…>
screen).
3.7.11.2
TEST EXECUTION
The last step of test initialization is to AutoRange the RPM4-E-DWT (see Section
3.7.11.1), then test execution begins.
The first step of test execution is to set the
E-DWT Lo Q-RPT shut off valve correctly
for the Q-RPT that will be used to run the
test (see 3.4). The RPM4-E-DWT display is
(
<XX>
is Lo or Hi):
Set valve for XX RPT
Set the E-DWT shut off valve to shut off or connect the RPM4-E-DWT’s Lo
Q-RPT as indicated by the E-DWT valve indicator LED (see Section 3.4). Turn
the valve knob fully towards the lit valve indicator LED until it reaches its stop.
Press
[ENT]
when ready to proceed with the test.
To abort the test but use the RPM4-E-DWT with the settings for the DUT that
was initialized, press [ESCAPE] from the <SET VALVE FOR xx RPT> screen. This can
be useful for setting up the RPM4-E-DWT to make DUT adjustments or run tests
other than the one defined by the AutoTest.
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...