![DH Instruments E-DWT-10000-AF Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/dh-instruments/e-dwt-10000-af/e-dwt-10000-af_operation-and-maintenance-manual_2492026012.webp)
E-DWT-10000-AF OPERATION AND MAINTENANCE MANUAL
© 2008 DH Instruments, a Fluke Company
Page 2
1.2
SPECIFICATIONS
1.2.1
GENERAL SPECIFICATIONS
Power requirements:
To RPM4-E-DWT:
To 12 VDC AC to DC power
supply :
12 VDC, 1.2 A
100 to 240VAC, 50-60 Hz
Temperature
Operating:
Storage:
18 to 28 °C
- 20 to 70 °C
Weight
E-DWT-10000-AF unit:
E-DWT system w/ accessories
in case:
14 kg (30 lb) approx.
27.3 kg (60 lb) approx.
Dimensions
E-DWT-10000-AF footprint:
E-DWT-10000-AF height:
System transport case:
41.4 cm W x 37.1 cm D (16.3 in. x 14.6 in.)
26.9 cm (10.6 in.), 33.6 cm (13.2 in.) to max variable volume handle stroke
39.4 cm H x 79.5 cm W x 51.8 cm D (15.5 in. x 31.3 in. x 20.4 in.)
Pressure range:
0 to 10 000 psi (70 MPa) gauge pressure
Operating medium:
Oil (di-ethyl-hexyl sebacate)
Reservoir capacity:
300 cc (18 in.
3
)
Variable volume displacement:
3 cc (0.18 in.
3
)
Priming pump displacement:
3.7 cc (0.23 in.
3
)
TEST pressure connection:
DH500 female, adaptors to 1/4 in. and 1/8 in. NPT female included.
DH500 is a gland and collar type fitting for 6mm (1/4 in.) coned and left
hand threaded tubes equivalent to AE F250C, HIP HF4, etc.
Pressure Limits:
Maximum Working
Pressure: 10 000 psi (70 MPa)
Maximum Pressure Without Damage: 12 000 psi (83 MPa)
Internal rupture disk burst pressure: 11500 psi (79 MPa)
When Lo Q-RPT is connected to TEST pressure:
Maximum Working
Pressure: 1 000 psi (7 MPa)
Maximum Pressure Without Damage: 2200 psi (15.2 MPa)
Pressure relief valve setting: 1 500 psi (10.3 MPa)
Maximum priming pump pressure: 100 psi (700 kPa)
Communication ports:
RS232 (COM1, COM2)
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...