3. OPERATION
Page 27
© 2008 DH Instruments, a Fluke Company
3.6.4
SOUNDS
RPM4-E-DWT is equipped with a variable frequency tone device to provide audible feedback
and alarms. The beeper is used for the following indications.
Valid key press
Brief beep. Choice between three frequencies or NO
sound is available (see Section 3.9.5.2).
Invalid key press
Descending two tone “blurp”
Change E-DWT Lo Q-RPT
shut-off valve setting
Four one second beeps (see Section 3.4).
Leak check completed
Three two second beeps (see Section 3.7.6).
Upper limit exceeded
Intermittent one second beeps (see Section 3.8.4).
Pmax! (overpressure limit)
exceeded
Eight second high frequency beep (see Section 3.8.4.1).
AutoTest in/out of tolerance
reading
Ascending triad/descending triad (see Section 3.7.11).
Hi Q-RPT is selected but Lo
Q-RPT is being pressurized
Intermittent two second high frequency beep. The audible
alarm is combined with flashing of the red,
Lo Q-RPT
Active
LED on the E-DWT (see Section 3.4, 3.7.2).
AutoTest in/out of tolerance
reading
Ascending triad/descending triad (see Section 3.7.11).
3.7
RPM4-E-DWT DIRECT FUNCTION KEYS
3.7.1
SUMMARY
Table 3 provides a brief summary of RPM4-E-DWT direct function key operation. It may be useful
to keep a copy of this summary near the E-DWT, especially when first becoming acquainted with its
operation.
Local operation of RPM4-E-DWT is through the front panel keypad. To minimize multi-
layered menu structures, the keypad numerical keys also provide direct access to the most
commonly used functions. The function accessed is labeled on the bottom half of the key.
Direct function keys are active whenever RPM4-E-DWT is in its MAIN RUN screen. Table 3
summarizes the operation of the direct function keys.
Table 3.
Summary of RPM4-E-DWT function key operation
DIRECT FUNCTION KEYS ARE ACTIVE FROM THE MAIN RUN SCREEN
See corresponding manual sections for full detail.
Adjust height of fluid head correction.
Set to zero to defeat correction.
View active range and toggle through available
ranges.
[ENT]
on a range activates it.
Feature not used in RPM4-E-DWT.
Select pressure unit of measure for the active
range. Choice of units available in this menu
can be customized.
Run the AutoZ function to AutoZero
the active Q-RPT.
Not used. E-DWT-AF RPM4-E-DWT is gauge
mode only.
Set up and optimize RPM4-E-DWT
features for a specific operating range.
Menu of less commonly used internal functions
and settings.
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...