3. OPERATION
Page 31
© 2008 DH Instruments, a Fluke Company
Table 4.
Settings made by AutoRange
SETTING/SELECTION
PURPOSE
SET TO
SECTION
Unit
Set pressure unit of measure
Operator specified during
AutoRange
3.7.3
Full scale
Establish span maximum
Operator specified during
AutoRange
None
AutoRange span
Basis of span based setting
calculations
Zero gauge to AutoRange full
scale
None
Q-RPT
Select most appropriate Q-RPT
to cover range and mode.
Q-RPT with lowest span whose
full scale is greater than
AutoRange full scale
1.2.3.1
Upper limit
Set maximum and minimum
pressure alarm limits.
105 % of operator entered full
scale pressure
3.8.4
Resolution
Set display resolution
0.01% of AutoRange span or
1 ppm of default Q-RPT span,
whichever is larger.
3.8.2
Stability Limit
Set Ready/Not Ready stability
criterion.
0.01% AutoRange span or 5
ppm of default Q-RPT span,
whichever is larger.
3.8.3,
3.7.5
OPERATION
To use the AutoRange function, press
[AutoRange]
and respond to the RPM4-E-DWT’s
prompts.
Select pressure unit of measure:
The screen is identical in appearance and function
to the
[UNIT]
screen (see Section 3.7.3).
Enter the AutoRange full scale pressure:
Enter the full scale pressure value in the
<------>
entry field and press
[ENT]
. If the full scale value entered cannot be covered by
RPM4-E-DWT, an error message appears.
View proposed AutoRange range:
The proposed AutoRange range screen is:
1. Indication that this is a screen of the proposed AutoRange
range.
2. Designator and position of the Q-RPT that AutoRange has
selected to cover the AutoRange full scale and measurement
mode.
3.
Pressure unit of measure specified in AutoRange.
4.
AutoRange full scale pressure.
AutoRange A70M Hi
psi 5000g
If AutoRange selects a Q-RPT but you would like to use the other one, the cursor control
keys may be used to toggle between Q-RPTs so that the other can be selected.
Accept proposed AutoRange:
To accept the proposed AutoRange and go to the MAIN
RUN screen with the new AutoRange range active, press
[ENT]
or
[AutoRange]
. To
modify the AutoRange full scale or pressure unit of measure, use
[ESC]
to back through
the AutoRange screens and make changes.
The E-DWT Lo Q-RPT Shut off valve protects the Lo Q-RPT (1 000 psi max) from
overpressure when the Hi Q-RPT is in use. Familiarize yourself with E-DWT operation (see Section
3.4) before changing ranges and applying pressures. Always turn the Lo Q-RPT Shut off valve fully
in the direction of the lighted green LED before applying pressure.
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...