5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS
Page 99
© 2008 DH Instruments, a Fluke Company
Use
[SPECIAL]
,
<8cal>
,
<1Hi Q-RPT >
or
<2Lo Q-RPT >
,
<1edit>
, to read and record
the current values of PA and PM for the Q-RPT to be calibrated.
AutoZ the RPM4-E-DWT with the reservoir shut off valve close, TEST shut off valve open
and open to atmosphere at the E-DWT TEST port.
Run the recommended calibration point sequence for the Q-RPT (see Section 5.3.4).
Record the pressure applied by the standard and the RPM4-E-DWT reading at each calibration
point. Dwell at least 60 seconds at each point after setting the reference pressure to allow full
stabilization.
Enter the calibration pressures and RPM4-E-DWT readings into a spreadsheet.
Calculate the “non-corrected” RPM4-E-DWT readings by backing out the PM recorded in
Steps above, following:
non-corrected reading = corrected reading/PM [Note: PA should be 0]
Perform a linear regression to find the offset and slope that best fit the non-corrected
RPM4-E-DWT readings to the calibration standard pressures. The offset is the new value of
PA, the slope is the new value of PM. If the RPM4-E-DWT was AutoZeroed properly to
calibration, HEAD was set to zero and the pressures from the calibration reference were
properly adjusted to the RPM4-E-DWT reference level (TEST port), the value of PA should
be very close to zero.
Press
[SPECIAL]
,
<8Cal>
,
<1Hi Q-RPT >
or
<2Lo Q-RPT >
,
<2edit>
and write the new
calibration date and the new values of PA and PM for the Q-RPT calibrated (see Section
5.3.6).
Calculate as left data for the calibration if desired following:
as left reading = non-corrected reading • new PM [Note: PA should be 0]
Perform additional verification pressure runs as desired.
5.4
E-DWT MAINTENANCE
There are no E-DWT maintenance procedures other than the overhaul procedure.
5.4.1
E-DWT OVERHAUL
E-DWT-10000-AF is generally overhauled as one common assembly. Removal of RPM4-E-DWT
from the E-DWT is not recommended unless required to perform RPM4-E-DWT and/or E-DWT
repairs. See Section 5.5.1 if it is necessary to remove the RPM4-E-DWT from the E-DWT.
Any or all of the following items may be included as part of an E-DWT-10000-AF overhaul:
•
Check that RPM4-E-DWT rear panel mounted cooling fan operates when RPM4-E-DWT
is powered.
•
Clean and inspect E-DWT TEST port connection.
•
Purge and clean reservoir and internal hydraulic circuit. Replace used oil with clean oil.
•
Check that internal fittings, screws, bolts and nuts are tight.
•
Perform
system
leak
and operational check.
•
Check Lo Q-RPT pressure relief valve for proper operation. It should crack at about
1 500 psi. Do not apply pressure greater than 1 500 psi to the RPM4-E-DWT Lo Q-RPT
as it may be damaged beyond repair. The RPM4-E-DWT warranty does not cover
damage due to Q-RPT overpressure.
•
Clean RPM4-E-DWT and E-DWT front panel.
•
Verify that RPM4-E-DWT internal barometer reads atmospheric pressure within
± 0.15 psi (1 kPa). Adjust if necessary (See Section 5.2.2).
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...