E-DWT-10000-AF OPERATION AND MAINTENANCE MANUAL
© 2008 DH Instruments, a Fluke Company
Page 46
3.8
[SETUP]
PURPOSE
[SETUP]
accesses a menu of functions and features commonly used in setting up and using the RPM4-E-DWT.
OPERATION
To access the SETUP menu, press
[SETUP]
from the MAIN RUN
screen. The display is:
1range 2res 3stab
4UL 5ATest
SETUP menu choices include:
<1range>
Save and delete ranges created by AutoRange (see Section 3.8.1).
<2res>
Adjust the resolution of pressure displays (see Section 3.8.2).
<3stab>
Adjust the pressure stability limit that is the criterion for the
Ready/Not Ready
indication
(see Section 3.8.3).
<4UL>
Adjust upper pressure limit alarm (see Section 3.8.4).
<5ATest>
Set up File AutoTest files and view data files resulting from running AutoTests (See
Section 3.8.5).
3.8.1
<1RANGE>
PURPOSE
Save AutoRange ranges and associated settings for recall. Delete previously saved
AutoRange ranges.
PRINCIPLE
The RPM4-E-DWT AutoRange function creates a range and automatically sets various
operating parameters proportionally to the range (see Section 3.7.5). Operating settings can
then be changed by the user. Many operating settings are specific to the current range (see
Table 2).
[SETUP]
,
<1range>
allows the AutoRange range and associated settings to be saved for
recall using the
[RANGE]
key (see Section 3.7.2). This can avoid having to recreate
frequently used ranges and settings.
[SETUP]
,
<1range>
is also used to delete ranges that have been saved but are no longer
needed.
3.8.1.1
SAVING AN A
UTO
R
ANGE
RANGE
OPERATION
To save a range created by AutoRange, the range must be the active range.
Use AutoRange or AutoTest to create the range (sees Sections 3.7.5, 3.7.11)
and then make any desired feature and setting adjustments.
Once the AutoRange range is set up as desired, press
[SETUP]
,
<1range>
,
<1save>
. The display is:
1.
Indication that this is a range to be saved.
2.
Designator and position of the Q-RPT used by
the range to be saved.
3.
Pressure unit of measure of the range to be
saved.
4.
Full scale pressure of the range to be saved.
Save range A70M Hi
psi 5000g
Содержание E-DWT-10000-AF
Страница 10: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 18: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 8 N NO OT TE ES S...
Страница 24: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...
Страница 102: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 92 N NO OT TE ES S...
Страница 128: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 118 N NO OT TE ES S...
Страница 130: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 120 N NO OT TE ES S...
Страница 132: ...E DWT 10000 AF OPERATION AND MAINTENANCE MANUAL 2008 DH Instruments a Fluke Company Page 122 N NO OT TE ES S...