HD31
- 134 -
V1.6
LP471UVC
Probe for the measurement of
UVC
irradiance
, with SICRAM module.
Measuring range (W/m
2
) 1x10
-3
…999.9x10
-3
1.000…19.999 20.00…199.99 200.0…1999.9
Resolution(W/m
2
) 0.1x10
-3
0.001 0.01 0.1
Spectral range
220 nm…280 nm (Peak 260 nm)
Calibration incertitude
<5%
f
2
(response as law of cosines)
<6%
f
3
(linearity)
<1%
f
4
(error in instrument reading )
±
1digit
f
5
(fatigue)
<0.5%
1 year drift
<2%
Operating temperature
0…50
°
C
Typical response curve of
LP471UVC
probe:
LP471BLUE
Probe for the measurement of
effective irradiance
in blue light spectral
band
, with SICRAM module.
Measuring range (W/m
2
) 0.1x10
-3
…999.9x10
-3
1.000…19.999 20.00…199.99 200.0…1999.9
Resolution(W/m
2
) 0.1x10
-3
0.001 0.01 0.1
Spectral range
380 nm…550 nm. action curve for blue-light induced damage B(
λ
)
Calibration incertitude
<10%
f
2
(response as law of cosines)
<6%
f
3
(linearity)
<3%
f
4
(error in instrument reading )
±
1digit
f
5
(fatigue)
<0.5%
1 year drift
<2%
Operating temperature
0…50
°
C
Typical response curve of
LP471BLUE
probe:
R
el
ati
ve
s
pe
ctr
al
r
esp
on
se
λ
(nm)
R
el
ati
ve
s
pe
ctr
al
r
esp
on
se
λ
(nm)