HD31
- 100 -
V1.6
13.5
C
ALIBRATING THE CO
2
PROBE
Three options are shown by selecting a CO
2
probe
for calibration:
•
ZERO/FRESH AIR
: calibrates the probe in one single point (any value included in the
measuring range).
•
LINEAR GAINS
: calibrates the probe in two, three or four points.
•
FACTORY
: sets user calibration equal to factory calibration. This is useful if wrong user
calibration data have been entered and you are temporarily unable to perform a new ca-
libration.
Fig. 13.5.1: calibrations for CO
2
probes
Note
: the value used for the barometric pressure compensation is shown at the bottom of the
display; if necessary, change the setting as indicated at paragraph 9.3 at page 74.
Select the desired type of calibration using
S
and
T
, then press the
F3
key (
OK
command) to
confirm.
If a message appears reminding that the operation will change the probe calibration data, se-
lect
YES
(
F3
key) to continue (or press F1 to select NO and exit).
13.5.1
ZERO/FRESH
AIR
OPTION FOR
CO
2
PROBES
By selecting the ZERO/FRESH AIR calibration option, the instrument displays the calibration
point (
SETPOINT
line) and the CO
2
value measured by the probe (
MEASURE
line). The calibra-
tion point can be adjusted using
S
and
T
.
Fig. 13.5.2: ZERO/FRESH AIR calibration for CO
2
probes
Calibration can be performed:
•
Positioning the probe in environment with known CO
2
concentration value, e.g.
in clean
air
.
Reference value
Measured value