HD31
- 41 -
V1.6
Fig. 4.6.3: connecting LP PYRA 06 albedometer
In order to calculate the global radiation in W/m
2
with respect to the output voltage signal of the
pyranometer, the pyranometer sensitivity must be set in the instrument, supplied with the pyra-
nometer itself. The sensitivity must be set in nV/(Wm
-2
) and range between 5000 and
30000 nV/(Wm
-2
). The default value is 10000 nV/(Wm
-2
). To set the sensitivity:
1.
Press
SELECT
to display the
MENU
command in the command bar, then select
MENU
by
pressing the
F2
key.
2.
Select the
CALIBRATION
item using
S
and
T
, then press
F3
(
OK
command) to confirm.
Note
: CALIBRATION item is not shown if the instrument is in protected mode. See para-
graph 8.2 to change the protection status.
3.
The list of the connected probes is displayed. Each probe is identified by its own serial num-
ber and by the type of measurement. Select VP472 module using
S
and
T
, then press
F3
(
OK
command) to confirm.
4.
Set the first digit of the pyranometer sensitivity which measures the
incident
radiation us-
ing
S
and
T
, then select the command "-->" (
F2
key) to proceed to the next digit (the cur-
rently selected digit flashes). To return to the previous digit, select "<--" (
F1
key). After
setting all the digits, press
OK
(
F3
key) to confirm the value.
5.
Set the first digit of the pyranometer sensitivity which measures the
reflected
radiation us-
ing
S
and
T
, then select the command "-->" (
F2
key) to proceed to the next digit (the cur-
rently selected digit flashes). To return to the previous digit, select "<--" (
F1
key). After
setting all the digits, press
OK
(
F3
key) to confirm the value.
Note
: the second sensitivity is necessary only if an albedometer is connected; if this isn't
the case, leave the default value.
6.
Press
ESC
twice to return to the measurement mode.
4.7
P
HOTOMETRIC AND RADIOMETRIC PROBES WITH
SICRAM
MODULE
The photometric and radiometric probes of
LP471...
series can be connected to the instru-
ment.
The measurement unit is chosen by the instrument depending on the probe connected. If sev-
eral measurement units are provided for the same probe, see paragraph 3.2 at page 13 for the
selection of the desired unit.
All the probes are factory-calibrated and do not require user calibrations.
Lower
terminal
Upper terminal block
LP PYRA 06
Red (+)
Blue (-)
Black (shield)
Green (+)
Brown (-)