HD31
- 117 -
V1.6
16
RESET OF THE INSTRUMENT
In case of blocked functionalities of the instrument, a hardware reset can be performed
through the button placed inside the battery compartment.
Proceed as follows:
1.
Remove the protective rubber shell, if any.
2.
Untighten the 4 screws fixing the battery compartment cover on the back of the instru-
ment.
3.
Press (e.g. using a small screwdriver) the reset button on top left of the battery.
Fig. 16.1: reset button
4.
Close the battery compartment with the 4 screws.
After reset, the instrument enters a waiting state for a new firmware for two minutes: press
ESC
to immediately exit the wait state and enter the normal measurement mode. If you ex-
ceed the two-minute time limit, the instrument switches off; in this case, press
ON/OFF
to
switch it on again: the instrument restarts in normal measuring mode.
The reset sets back instrument parameters to factory defaults.
After a reset operation,
check configuration and, if necessary, reset operating parameters to the desired values.
Reset button