HD31
- 56 -
V1.6
3.
Using
S
and
T
, select
START
to set the start time, or
STOP
command to set the stop time. If a programmed time has not
been entered, the fields appear empty, otherwise the entered
time is shown. Select
OK
by pressing the
F3
key to continue.
4.
Select the different date/time fields using “
<<<
” (
F1
key) or
“
>>>
” (
F2
key) commands, the selected field flashes. Use
S
and
T
to change the value.
5.
Press
F3
(OK command) to confirm.
6.
Press
ESC
to return to measuring mode. The programmed
logging symbol appears in the display status bar.
At the set start date/time, the logging of the measurement lines
explicitly selected for logging, or the logging of all the meas-
urement lines if no line was explicitly selected for logging is
started. The LOG indication, the set logging interval and the
logging duration are shown in the display status bar.
Logging will automatically stop at the set stop date/time. If you want logging to be stopped in
advance, date/time programming must be deleted and stop logging manually as explained be-
low:
1.
Press
SELECT
to display the
LOG
command in the command bar, then select the command
by pressing the
F1
key.
2.
Select the
PROG
command by pressing the
F2
key.
3.
Select the
CLR
command by pressing the
F1
key. A message
is displayed asking to confirm the operation, select
YES
(
F3
key) to continue: logging programming is deleted.
4.
Press
ESC
to return to the previous command level.
Programmed logging