HD31
- 131 -
V1.6
20.1.6
Photometry and radiometry
LP471PHOT
Probe for the measurement of
illuminance
, with SICRAM module.
Measuring range (lux)
0.10…199.99 …1999.9 …19999 …199.99x10
3
Resolution(lux)
0.01
0.1
1
0.01 x 10
3
Spectral range
in accordance with standard photopic curve V(
λ
)
α
(temperature coefficient) f
6
(T) <0.05%
K
Calibration incertitude
<4%
f'
1
(accordance with photopic response
V(
λ
))
<6%
f
2
(response as law of cosines)
<3%
f
3
(linearity)
<1%
f
4
(error in instrument reading)
<0.5%
f
5
(fatigue)
<0.5%
Class B
1 year drift
<1%
Operating temperature
0…50
°
C
Reference standard
CIE n°69 – UNI 11142
LP471LUM2
Probe for the measurement of
luminance
, with SICRAM module.
Measuring range (cd/m
2
) 0.1…1999.9
…19999 …199.99x10
3
…1999.9x10
3
Resolution(cd/m
2
) 0.1
1
0.01 x 10
3
0.1 x 10
3
Angle of view
2
°
Spectral range
in accordance with standard photopic curve V(
λ
)
α
(temperature coefficient) f
6
(T) <0.05%
K
Calibration incertitude
<5%
f'
1
(accordance with photopic response
V(
λ
)
<8%
f
3
(linearity)
<1%
f
4
(error in instrument reading)
<0.5%
f
5
(fatigue)
<0.5%
Class C
1 year drift
<1%
Operating temperature
0…50
°
C
Reference standard
CIE n°69 – UNI 11142
Typical response curve of
LP471PHOT
and
LP471LUM2
probes:
R
el
ati
ve
s
pe
ctr
al
r
esp
on
se
λ
(nm)