HD31
- 55 -
V1.6
7.3
M
ANUALLY START
/
STOP AUTOMATIC LOGGING
1.
Press
SELECT
to display the
LOG
command in the command bar,
then select the command by press-
ing the
F1
key.
2.
To start logging, select the
RUN
command by pressing the
F1
key.
Note
: the RUN command is replaced by the STOP command if
logging is in progress.
Note
: the RUN command will not go into effect if the instru-
ment is waiting for a programmed startup.
3.
The logging of the measurement lines selected for logging, or
the logging of all the measurement lines if no line was explic-
itly selected for logging, is started. The LOG indication, the
set logging interval and the logging duration appear in the
display status bar.
4.
To stop logging, select again the
LOG
command (
F1
key),
then select the
STOP
command (
F1
key).
5.
After the stop command, the logging stop message appears: wait for the completion of the
operation.
7.4
P
ROGRAMMING AUTOMATIC LOGGING START
/
STOP TIME
1.
Press
SELECT
to display the
LOG
command in the command bar,
then select the command by press-
ing the
F1
key.
2.
Select the
PROG
command by pressing the
F3
key to pro-
gram the logging start/stop time.