HD31
- 57 -
V1.6
5.
Select the
STOP
command by pressing the
F1
key.
Note
: if you try to stop logging manually without first deleting
the programming function, the instrument displays an error
message ("Log: Event conflict").
Only the logging start time can be set, in this case it will have to be stopped manually, or only
the logging stop time, in this case it will have to be started manually.
After programming the logging function, the instrument can be switched off: it will be switched
on automatically at the date/time set for the logging startup.
7.5
A
UTO
-
OFF DURING AUTOMATIC LOGGING
If the logging interval is lower than 60 seconds (120 seconds if the CO
2
probe is connected),
the instrument will remain switched on during logging. If the logging interval exceeds or is
equal to 60 seconds (120 seconds if the CO
2
probe is connected), the instrument can be set to
switch off between to subsequent loggings, to save battery life. If Auto-off is enabled, the in-
strument will automatically switch on at sampling and switch off immediately after. To enable
and disable Auto-off:
1.
Press
SELECT
to display the
MENU
command in the command bar, then
select the command by pressing the
F2
key.
2.
Select the
LOG SETTINGS
item using
S
and
T
, then press
the
F3
key (
OK
command) to confirm.
Note
: the item is disabled if logging is in progress.
3.
Select the
SLEEP
item using
S
and
T
, then press the
F3
key
(
OK
command) to confirm.