4.1.1
S
ETTING THE INSTRUMENT TO READ A
PRT
PROBE WITHOUT
SICRAM
MODULE
If a PRT probe without SICRAM module is connected, the type of connected probe must be set
in the instrument:
1.
Turn the instrument off and connect the probe.
2.
Turn the instrument on.
3.
Press
SELECT
to display the
MENU
command in the command bar, then
select the command by pressing
F2
.
4.
Select the
SETUP
item using
S
and
T
, then press
F3
(
OK
command) to confirm.
5.
Select the
CHANNELS
item using
S
and
T
, then press
F3
(
OK
command) to confirm.
6.
Using
S
and
T
, select the input (A, B or C) of the instru-
ment to which the probe is connected, then press
F3
(
OK
command) to confirm.