HD31
- 132 -
V1.6
LP471PAR
Quantum-radiometric probe for the measurement of
photon flux in the PAR
chlorophyll field
, with SICRAM module.
Measuring range (
μ
mol/m
2
s)
0.01… 199.99
200.0…1999.9 2000…10000
Resolution(
μ
mol/m
2
s)
0.01
0.1 1
Spectral range
400 nm…700 nm
Calibration incertitude
<5%
f
2
(response as law of cosines)
<6%
f
3
(linearity)
<1%
f
4
(error in instrument reading)
±
1digit
f
5
(fatigue)
<0.5%
1 year drift
<1%
Operating temperature
0…50
°
C
Typical response curve of probe
LP471PAR
:
LP471RAD
Probe for the measurement of
irradiance
, with SICRAM module.
Measuring range (W/m
2
) 0.1x10
-3
…999.9x10
-3
1.000…19.999 20.00…199.99 200.0…1999.9
Resolution(W/m
2
) 0.1x10
-3
0.001 0.01 0.1
Spectral range
400 nm…1050 nm
Calibration incertitude
<5%
f
2
(response as law of cosines)
<6%
f
3
(linearity)
<1%
f
4
(error in instrument reading )
±
1digit
f
5
(fatigue)
<0.5%
1 year drift
<1%
Operating temperature
0…50
°
C
Typical response curve of
LP471RAD
probe:
R
el
ati
ve
s
pe
ctr
al
r
esp
on
se
λ
(nm)
R
el
ati
ve
s
pe
ctr
al
r
esp
on
se
λ
(nm)