PROTECTED MODE
The calibration of the connected probes can be protected against unwanted alterations by en-
tering a security password. The password (8 digits) is factory-set to the value
11111111
(eight times 1) and can be changed.
Attention: if the password is changed, be sure to keep it in a safe place.
In case of loss
of the password, you should contact an authorized Technical Assistance Service.
Note
: the password setup is not available if the instrument is in logging; in that case, stop the
logging in progress to proceed with configuration.
8.1
P
ASSWORD CHANGE FOR THE PROTECTED MODE
For security reasons, the preset default password should be changed. To change the password,
proceed as follows:
1.
Press
SELECT
to display the
MENU
command in the command bar,
then select the command by press-
ing the
F2
key.
2.
Select the
LOCKED CONFIG
item by using
S
and
T
, then
press the
F3
key (
OK
command) to confirm.
3.
Select the
PSWD CHANGE
item by using
S
and
T
, then
press the
F3
key (
OK
command) to confirm.
Note
: PSWD CHANGE item is not displayed if the instrument
is in secure mode. See paragraph 8.2 to change the protec-
tion status.