EZT-570i Technical Manual
31
5. Remote PC Communication Problems
This section covers troubleshooting basics for user communications to the EZT. This is for PC to
EZT communications, not internal communication between EZT components. The user
communication capabilities are enabled from the “Web Server/Modbus/VNC” settings screen under
the chamber setup menu.
Figure 5-1 Web Server/Modbus/VNC Communication Settings
For more detailed information on communication format, commands and functionality of the
232/485 serial interfaces, see the EZT-570i User Communication Reference Manual.
When connecting remote devices to the EZT’s communication ports, make sure that
the equipment is properly grounded as required by the manufacturer’s instructions
with a good earth ground. Poor site earths can introduce electrical noise and transient
voltages into the communication wiring resulting in poor performance or damage to
equipment.
Содержание EZT-570i
Страница 1: ...EZT570i Technical Manual Rev B EZT 570i Technical Manual...
Страница 14: ...EZT 570i Technical Manual 14 4 1 6 Conditioning System Logic Flow...
Страница 15: ...EZT 570i Technical Manual 15...
Страница 19: ...EZT 570i Technical Manual 19 4 2 2 Humidity System Logic Flow...
Страница 21: ...EZT 570i Technical Manual 21 4 3 2 Auxiliary Cooling System Logic Flow...
Страница 23: ...EZT 570i Technical Manual 23 4 4 2 Dry Air Purge System Logic Flow...
Страница 25: ...EZT 570i Technical Manual 25 4 5 2 Altitude System Logic Flow...
Страница 29: ...EZT 570i Technical Manual 29 4 7 2 Transfer System Logic Flow...
Страница 30: ...EZT 570i Technical Manual 30...
Страница 58: ...EZT 570i Technical Manual 58 7 2 Standard Output Configuration...
Страница 92: ......
Страница 93: ...EZT 570i Technical Manual 1 Appendix...