EZT-570i Technical Manual
84
8.9 Configuration
Options
The “Configuration Options” screen is accessed from the “Chamber” menu. This screen is used to
set general values relating to altitude system operation and other options.
Figure 8-14 Altitude Options
Set the value of each item by touching the respective number field. Enter the correct value using the
keypad and press done. Enable the digital inputs or special settings by pressing the respective
enable buttons.
Parameter Descriptions
Window Heater Enable SP:
Sets the temperature below which the window heater turns on.
Alt Hi Speed Fan Enable SP:
Sets the altitude (in Kft) above which the high speed fans will be
turned on.
Alt Cond System Disable:
Sets the altitude above which the conditioning system will be
automatically disabled.
Digital Inputs Off/On:
Enables/disables the digital input option on the EZT. The digital
input card must be installed on the EZT for the option to work.
Special Settings Off/On:
Enables/disables the special settings screen in the EZT application
for custom settings use. If used on a chamber, it will be application
specific.
Содержание EZT-570i
Страница 1: ...EZT570i Technical Manual Rev B EZT 570i Technical Manual...
Страница 14: ...EZT 570i Technical Manual 14 4 1 6 Conditioning System Logic Flow...
Страница 15: ...EZT 570i Technical Manual 15...
Страница 19: ...EZT 570i Technical Manual 19 4 2 2 Humidity System Logic Flow...
Страница 21: ...EZT 570i Technical Manual 21 4 3 2 Auxiliary Cooling System Logic Flow...
Страница 23: ...EZT 570i Technical Manual 23 4 4 2 Dry Air Purge System Logic Flow...
Страница 25: ...EZT 570i Technical Manual 25 4 5 2 Altitude System Logic Flow...
Страница 29: ...EZT 570i Technical Manual 29 4 7 2 Transfer System Logic Flow...
Страница 30: ...EZT 570i Technical Manual 30...
Страница 58: ...EZT 570i Technical Manual 58 7 2 Standard Output Configuration...
Страница 92: ......
Страница 93: ...EZT 570i Technical Manual 1 Appendix...