EZT-570i Technical Manual
77
Do not change the refrigeration type to any selection other than what is installed on
the chamber and called out on the EZT Configuration form supplied with the chamber
or the refrigeration system may not operate properly and damage may occur.
Cmp On %:
Sets the loop percentage of output required to turn enable the refrigeration
system because cooling is required.
Cmp On Dly:
Sets the delay in seconds for which the loop output must exceed the cmp on %
before the refrigeration system is turned on.
Cmp Off %:
Sets the loop percentage of output required to disable the refrigeration system
because cooling is no longer required.
Cmp Off Dly:
Sets the delay in seconds for which the loop output must exceed the cmp off %
before the refrigeration system is turned off.
Stager Dly:
Sets the stager start delay time in seconds for system 2 on cascade systems.
Defrost:
Sets the defrost mode based on installed defrost components.
-
“Defrost
On”
is
used
for
smaller compressors (typically smaller than 7.5hp)
-
“Large
HP
Defrost”
is
used
for
larger systems (typically larger than 7.5hp)
The large horsepower defrost is used to control the cascade condenser liquid line
cooling solenoids on cascade systems. It cycles the solenoids on and off to
maintain system 2 head pressures rather than cycling the system 1 compressor
on and off. Cycling large horsepower compressors is detrimental to the life of the
compressor and is not recommended by the manufacturer.
The “
alternating
” defrost selections are only used for redundant refrigeration
systems that use independent conditioning plenums. This allows one system to
go into defrost while the other system conditions the chamber. The “Ind Plenum”
(independent plenums) selection under the advanced settings must be selected
prior to selecting an alternating defrost mode or the selection will not be
accepted.
Pumpdown On/Off:
Enables or disables pumpdown mode.
PumpDown Reset Time Out:
Sets the time in seconds for which the chamber can be without
power before pumpdown will be disabled and require manual
reset or the four hour automatic reset delay.
Содержание EZT-570i
Страница 1: ...EZT570i Technical Manual Rev B EZT 570i Technical Manual...
Страница 14: ...EZT 570i Technical Manual 14 4 1 6 Conditioning System Logic Flow...
Страница 15: ...EZT 570i Technical Manual 15...
Страница 19: ...EZT 570i Technical Manual 19 4 2 2 Humidity System Logic Flow...
Страница 21: ...EZT 570i Technical Manual 21 4 3 2 Auxiliary Cooling System Logic Flow...
Страница 23: ...EZT 570i Technical Manual 23 4 4 2 Dry Air Purge System Logic Flow...
Страница 25: ...EZT 570i Technical Manual 25 4 5 2 Altitude System Logic Flow...
Страница 29: ...EZT 570i Technical Manual 29 4 7 2 Transfer System Logic Flow...
Страница 30: ...EZT 570i Technical Manual 30...
Страница 58: ...EZT 570i Technical Manual 58 7 2 Standard Output Configuration...
Страница 92: ......
Страница 93: ...EZT 570i Technical Manual 1 Appendix...