EZT-570i Technical Manual
81
Hum Max SP:
Sets the maximum air temperature set point at which the humidity system will
operate. When the air temperature set point and/or air temperature is above this
value, the humidity system will be automatically turned off and the system status
indicator “RH TMP DISABLE” will be lit.
Temp Comp On/Off:
Enables or disables the humidity sensor temperature compensation
algorithm in the EZT depending on whether or not a compensated
sensor is used.
The compensation is for a Vaisala HMM30C sensor only. Do not
use it for other brands of sensors. This should be disabled when
using compensated humidity sensors. If set incorrectly, the EZT
will not read the correct humidity in the chamber.
Condensation On/Off:
Enables or disables the condensation control option. This feature can
reduce the formation of condensation during testing on product within the
chamber. However, it can greatly reduce the performance of the
chamber. This option is used for special circumstances only.
Содержание EZT-570i
Страница 1: ...EZT570i Technical Manual Rev B EZT 570i Technical Manual...
Страница 14: ...EZT 570i Technical Manual 14 4 1 6 Conditioning System Logic Flow...
Страница 15: ...EZT 570i Technical Manual 15...
Страница 19: ...EZT 570i Technical Manual 19 4 2 2 Humidity System Logic Flow...
Страница 21: ...EZT 570i Technical Manual 21 4 3 2 Auxiliary Cooling System Logic Flow...
Страница 23: ...EZT 570i Technical Manual 23 4 4 2 Dry Air Purge System Logic Flow...
Страница 25: ...EZT 570i Technical Manual 25 4 5 2 Altitude System Logic Flow...
Страница 29: ...EZT 570i Technical Manual 29 4 7 2 Transfer System Logic Flow...
Страница 30: ...EZT 570i Technical Manual 30...
Страница 58: ...EZT 570i Technical Manual 58 7 2 Standard Output Configuration...
Страница 92: ......
Страница 93: ...EZT 570i Technical Manual 1 Appendix...