8.7 Purge/Lo RH Options
The “Purge/Lo RH Options” screen is accessed from the “Chamber” menu. This screen is used to
define what type of purge system is installed on the chamber and what dewpoint limits the humidity
system is allowed to control to, if equipped.
Figure 8-12 Purge/Lo RH Options
You must press the “Set” button to save changes to the purge system type selection. Be
sure to press the “Set” button prior to switching to another configuration screen or any
change to the purge system type will not be saved.
Parameter Descriptions
Purge/Low RH Type:
Sets the type of purge system installed.
- “Dry Air Purge” is used when just the air dryer is installed.
- “Low RH Option Installed” is used when the frozen coil option is installed.
Min Dewpoint Limit:
Sets the minimum dewpoint that the chamber will control to. In essence, this
limits the minimum humidity level that the chamber can achieve at a given
temperature.
This value is used to protect the refrigeration system from liquid refrigerant
flood back that can happen when trying to achieve to low of a humidity level
at lower air temperatures.
Standard humidity is limited to a 2C (35F) dewpoint.
The low RH option is limited to a -30.0C (-22F) dewpoint.
Max Dewpoint Limit:
Sets the maximum dewpoint that the chamber will control to. In essence, this
limits the maximum humidity that the chamber can achieve at a given
temperature.
This value is primarily used to protect modular walk-ins. The modular walk-in
panels can not withstand high temperature and humidity levels.
Standard
chambers
typically have a setting of 100C (212F).
Modular walk-ins are typically limited to 70C (158F).
Содержание EZT-570i
Страница 1: ...EZT570i Technical Manual Rev B EZT 570i Technical Manual...
Страница 14: ...EZT 570i Technical Manual 14 4 1 6 Conditioning System Logic Flow...
Страница 15: ...EZT 570i Technical Manual 15...
Страница 19: ...EZT 570i Technical Manual 19 4 2 2 Humidity System Logic Flow...
Страница 21: ...EZT 570i Technical Manual 21 4 3 2 Auxiliary Cooling System Logic Flow...
Страница 23: ...EZT 570i Technical Manual 23 4 4 2 Dry Air Purge System Logic Flow...
Страница 25: ...EZT 570i Technical Manual 25 4 5 2 Altitude System Logic Flow...
Страница 29: ...EZT 570i Technical Manual 29 4 7 2 Transfer System Logic Flow...
Страница 30: ...EZT 570i Technical Manual 30...
Страница 58: ...EZT 570i Technical Manual 58 7 2 Standard Output Configuration...
Страница 92: ......
Страница 93: ...EZT 570i Technical Manual 1 Appendix...