7.2.1.2 Custom Output Configuration (DTS, VTS, TSB)
Hot (left) Chamber (Q40):
This output is used to turn on the left chamber of a DTS or the
hot chamber of a VTS using chamber event 1. It is also used to
enable both baths for a TSB. When used for a DTS or VTS, the
output will turn off during a transfer to stop the fans and then turn
back on once the transfer is complete.
Cold (center) Chamber (Q2):
This output is used to turn on the center chamber of a DTS or
the cold chamber of a VTS using chamber event 2. It is not used
for a TSB. The output will turn off during a transfer to stop the
fans and then turn back on once the transfer is complete.
Hot (right) Chamber (Q44):
This output is used to turn on the right chamber of a DTS using
chamber event 7. The output will turn off during a transfer to
stop the fans and then turn back on once the transfer is
complete.
Maximum Heat (right) Enable (Q35):
This output is used to enable and disable the maximum heating
SSR for the right chamber of a DTS. When the loop output
exceeds the maximum output on percentage for the on delay
period set in the configurator, this output will turn on.
In Transfer Door Lock (Q45):
This output is used to enable a locking mechanism on the
chamber door (if equipped) to prevent the door from being
opened during a transfer. When the transfer is complete, the
output turns off.
Transfer Left (Q32):
This output is used to turn on the transfer mechanism to move
the basket to the left on a DTS or TSB. For a DTS it transfers
the left basket into the left (hot) chamber and the right basket
into the center (cold) chamber. On a TSB, it transfers the basket
from the cold or unload position to the unload or hot position.
Transfer Right (Q33):
This output is used to turn on the transfer mechanism to move
the basket to the right on a DTS or TSB. For a DTS it transfers
the right basket into the right (hot) chamber and the left basket
into the center (cold) chamber. On a TSB, it transfers the basket
from the hot or unload position to the unload or cold position.
Transfer Up (Q46):
This output is used to turn on the transfer mechanism to move
the basket up on a VTS or TSB equipped with a pneumatic
transfer mechanism. For a VTS, the basket is transferred into
the hot chamber. On pneumatic TSB systems, this lifts the
basket out of the baths so that it can move from side to side. It
also holds the basket up when the unload position is selected.
Transfer Down (Q47):
This output is used to turn on the transfer mechanism to move
the basket down on a VTS or TSB equipped with a pneumatic
transfer mechanism. For a VTS, the basket is moved into the
cold chamber. On pneumatic TSB systems, this lowers the
basket into the baths once it is positioned over the hot or cold
bath.
Содержание EZT-570i
Страница 1: ...EZT570i Technical Manual Rev B EZT 570i Technical Manual...
Страница 14: ...EZT 570i Technical Manual 14 4 1 6 Conditioning System Logic Flow...
Страница 15: ...EZT 570i Technical Manual 15...
Страница 19: ...EZT 570i Technical Manual 19 4 2 2 Humidity System Logic Flow...
Страница 21: ...EZT 570i Technical Manual 21 4 3 2 Auxiliary Cooling System Logic Flow...
Страница 23: ...EZT 570i Technical Manual 23 4 4 2 Dry Air Purge System Logic Flow...
Страница 25: ...EZT 570i Technical Manual 25 4 5 2 Altitude System Logic Flow...
Страница 29: ...EZT 570i Technical Manual 29 4 7 2 Transfer System Logic Flow...
Страница 30: ...EZT 570i Technical Manual 30...
Страница 58: ...EZT 570i Technical Manual 58 7 2 Standard Output Configuration...
Страница 92: ......
Страница 93: ...EZT 570i Technical Manual 1 Appendix...