EZT-570i Technical Manual
72
8.3 Loops/Monitor
Tagnames
The “Loops/Monitor Tagnames” screen is accessed from the “Loops” menu. This screen allows you
to define what type of control loop it is, what its name is and specific control output settings for the
loop.
Figure 8-7 Loop/Monitor Tagnames
The arrow keys provided on the upper right of the screen allow you to scroll through all available
control loops and monitor loops that have been set on the ”Number of Loops/Monitors” screen.
You must press the “Set” button to save changes to the input type. Be sure to press the “Set”
button prior to switching to another loop or another configuration screen or any change to the
input type will not be saved.
Parameter Descriptions
Input Type:
Sets the type of control loop (temperature, humidity, altitude, etc.).
Do not alter the loop type to any selection other than what is
called out on the EZT Configuration form supplied with the
chamber or the chamber may not operate properly and damage to
equipment can occur.
Tagname:
Sets the name of the loop (what its controlling). This description is used
throughout the loop and trend screens as well as input selection screen for data
logging in the main EZT-570i runtime application.
Alm1 Message:
Sets the alarm message that will be used for input alarms. This will be the name
displayed on the alarm monitor screen when the user configurable alarm is
active.
Eng Units:
Sets the units for the input (degrees C or F, RH, KFT, etc.). The units are
displayed on the single and dual loop display screen.
Dec Pos:
Sets how many decimal positions to use when displaying an input (always 1).
Max Heat%:
Sets the loop percentage of output at which the maximum heat output is enabled.
Содержание EZT-570i
Страница 1: ...EZT570i Technical Manual Rev B EZT 570i Technical Manual...
Страница 14: ...EZT 570i Technical Manual 14 4 1 6 Conditioning System Logic Flow...
Страница 15: ...EZT 570i Technical Manual 15...
Страница 19: ...EZT 570i Technical Manual 19 4 2 2 Humidity System Logic Flow...
Страница 21: ...EZT 570i Technical Manual 21 4 3 2 Auxiliary Cooling System Logic Flow...
Страница 23: ...EZT 570i Technical Manual 23 4 4 2 Dry Air Purge System Logic Flow...
Страница 25: ...EZT 570i Technical Manual 25 4 5 2 Altitude System Logic Flow...
Страница 29: ...EZT 570i Technical Manual 29 4 7 2 Transfer System Logic Flow...
Страница 30: ...EZT 570i Technical Manual 30...
Страница 58: ...EZT 570i Technical Manual 58 7 2 Standard Output Configuration...
Страница 92: ......
Страница 93: ...EZT 570i Technical Manual 1 Appendix...