EZT-570i Technical Manual
13
SYMPTOM PROBABLE
CAUSES
CORRECTIVE
ACTIONS
Compressor internal thermal overload
tripped.
Chamber in humidity mode.
Refrigeration system not enabled or
stager start delay time not met.
Allow compressor to cool. Check
refrigeration system/injection valve
operation.
Check humidity system type. System
2 disabled for single stage humidity
operation.
Check configurator settings.
Refrigeration system type should
match installed system type. Wait for
stager start delay period.
Chamber not cooling.
(further diagnosis and/or repair
requires certified refrigeration service
personnel)
Chamber event not on.
Conditioning system disabled (altitude
chambers).
Cooling coil fouled with ice build-up.
9300 controller output off.
Chamber in defrost (if equipped).
Refrigeration system capacity
exceeded.
Turn on chamber event.
Altitude above controllable limit for
temperature. Decrease altitude or
turn off altitude system.
Initiate defrost or warm up chamber to
melt ice from coil. Seal ports or leaks
in chamber to minimize moisture
migration into chamber and
accumulating on coil.
Verify proper 9300 controller
configuration. Check set point.
Replace 9300 controller.
Check defrost status. Allow defrost to
complete or terminate defrost.
Reduce live load in chamber.
Defrost not starting.
Suction line not below defrost
thermostat setting.
Defrost thermostat not working.
Check thermostat setting. Defrost not
required.
Check thermostat set point and
operation. Adjust/replace.
Defrost not terminating.
Defrost delay off period (15 minutes) not
completed.
Defrost thermostat not working.
Allow enough time for defrost
sequence to complete.
Check thermostat set point and
operation. Adjust/replace.
Содержание EZT-570i
Страница 1: ...EZT570i Technical Manual Rev B EZT 570i Technical Manual...
Страница 14: ...EZT 570i Technical Manual 14 4 1 6 Conditioning System Logic Flow...
Страница 15: ...EZT 570i Technical Manual 15...
Страница 19: ...EZT 570i Technical Manual 19 4 2 2 Humidity System Logic Flow...
Страница 21: ...EZT 570i Technical Manual 21 4 3 2 Auxiliary Cooling System Logic Flow...
Страница 23: ...EZT 570i Technical Manual 23 4 4 2 Dry Air Purge System Logic Flow...
Страница 25: ...EZT 570i Technical Manual 25 4 5 2 Altitude System Logic Flow...
Страница 29: ...EZT 570i Technical Manual 29 4 7 2 Transfer System Logic Flow...
Страница 30: ...EZT 570i Technical Manual 30...
Страница 58: ...EZT 570i Technical Manual 58 7 2 Standard Output Configuration...
Страница 92: ......
Страница 93: ...EZT 570i Technical Manual 1 Appendix...