EZT-570i Technical Manual
74
8.4 Chamber
Options
The “Chamber Options” screen is accessed from the “Chamber” menu. This screen is used to define
what type of chamber the EZT-570i is being used on and enables special logic for certain types of
chambers.
Figure 8-8 Chamber Options
Parameter Descriptions
All Options Off:
Used for all standard temperature/humidity/altitude chambers.
RC Logic Enabled:
Enables the logic for the RC blower motor. The logic ties event 8 to output
Q44 to turn the RC blower motor on and off with the event. Note that this
option is typically only used for retrofitting RC chambers that require event
control for the blower.
LC Logic Enabled:
Enables the logic for pump control and fluid system monitoring. The logic
turns on output Q44 for the pump with event 1 and enables the input alarm
logic for high temperature (I44), pressure (I45) and flow (I46) monitoring.
VTS Logic Enabled:
Enables the logic for basket position and dual zone control. Loop 1 is set for
hot chamber control and loop 2 for cold chamber control. The logic turns on
output Q44 for the hot chamber using event 1 while event 2 becomes the
cold chamber. Outputs Q46 and Q47 for basket up and down are enabled
through events 5 and 6.
Inputs I30 and I31 are used for basket position monitoring for hot and cold
chamber positions respectively. The door switch input (I42) is monitored to
disable the transfer if either door is opened.
DTS Logic Enabled:
Enables the logic for basket position and three zone control. Loop 1 is set for
the left (hot) chamber control, loop 2 for the center (cold) chamber control
and loop 3 is for the right (hot) chamber control. Loops 4 and 5 are set for
product control for the left and right basket respectively.
Содержание EZT-570i
Страница 1: ...EZT570i Technical Manual Rev B EZT 570i Technical Manual...
Страница 14: ...EZT 570i Technical Manual 14 4 1 6 Conditioning System Logic Flow...
Страница 15: ...EZT 570i Technical Manual 15...
Страница 19: ...EZT 570i Technical Manual 19 4 2 2 Humidity System Logic Flow...
Страница 21: ...EZT 570i Technical Manual 21 4 3 2 Auxiliary Cooling System Logic Flow...
Страница 23: ...EZT 570i Technical Manual 23 4 4 2 Dry Air Purge System Logic Flow...
Страница 25: ...EZT 570i Technical Manual 25 4 5 2 Altitude System Logic Flow...
Страница 29: ...EZT 570i Technical Manual 29 4 7 2 Transfer System Logic Flow...
Страница 30: ...EZT 570i Technical Manual 30...
Страница 58: ...EZT 570i Technical Manual 58 7 2 Standard Output Configuration...
Страница 92: ......
Страница 93: ...EZT 570i Technical Manual 1 Appendix...