System ‘B’ Boiler Thermostat (I46):
The system ‘B’ boiler thermostat input is only used when there
are independent system plenums since this usually results in a
boiler on each plenum. Thus, this input is used to indicate that
the boiler on plenum ‘B’ has overheated. This input is active low,
i.e., the input must be off in order for the alarm to be indicated.
When active (input off), the EZT shuts down the chamber until
the alarm is acknowledged (silence button pressed on alarm
monitor screen) and the input is on (requires manual reset of
thermostat).