EZT-570i Technical Manual
68
8. Adjusting EZT Configuration options
The EZT configurator is used to set the functionality of the EZT to match the chamber type and
available options on the chamber. The settings contained in the configurator affect the control of
chamber components and systems. Changes to these parameters should only be performed by or
under the supervision of CSZ authorized service personnel.
Changing certain configurator settings to values other than those listed on the EZT
Configuration sheet provided with the chamber can cause damage to the chamber
and present hazards to personnel. Changes to any configurator settings not
specifically called out in this manual may only be done with the consent of an
authorized CSZ representative.
Failure to comply with this WARNING will void the chamber’s warranty.
8.1 Accessing the EZT Configurator
To access the EZT configurator, you must quit the EZT570i application and start the configurator
program. To do this, you must have the proper security access to exit the EZT application. All
chambers leave the factory with a default user name and password. Select “Logon” under the
security menu and enter the default password information:
User: FACTORY
Password: CONTROL
Once you are logged in as “Factory”, you have the ability to exit the EZT570i application. To exit the
application, select “System Maintenance” from the chamber setup menu. Select “Exit Application”
from the system maintenance setup menu. Press the “Exit Application” button to quit the EZT-570i
application.
Figure 8-1 Exit Application
If security is not enabled, you will not have to log on to gain access to quit the application. However,
if security is enabled, and the default password is no longer valid, then you must either obtain the
necessary user name and password from the end user of the chamber, or you can bypass the
security by turning off power to the chamber.
Содержание EZT-570i
Страница 1: ...EZT570i Technical Manual Rev B EZT 570i Technical Manual...
Страница 14: ...EZT 570i Technical Manual 14 4 1 6 Conditioning System Logic Flow...
Страница 15: ...EZT 570i Technical Manual 15...
Страница 19: ...EZT 570i Technical Manual 19 4 2 2 Humidity System Logic Flow...
Страница 21: ...EZT 570i Technical Manual 21 4 3 2 Auxiliary Cooling System Logic Flow...
Страница 23: ...EZT 570i Technical Manual 23 4 4 2 Dry Air Purge System Logic Flow...
Страница 25: ...EZT 570i Technical Manual 25 4 5 2 Altitude System Logic Flow...
Страница 29: ...EZT 570i Technical Manual 29 4 7 2 Transfer System Logic Flow...
Страница 30: ...EZT 570i Technical Manual 30...
Страница 58: ...EZT 570i Technical Manual 58 7 2 Standard Output Configuration...
Страница 92: ......
Страница 93: ...EZT 570i Technical Manual 1 Appendix...