![Cryo-con 24C Скачать руководство пользователя страница 90](http://html.mh-extra.com/html/cryo-con/24c/24c_user-manual_2691809090.webp)
Cryo-con Model 24C
Basic Setup and Operation
The autotune sequence is initiated by selecting the Go field. If the initialization of
process modeling is successful, the status display line will change from idle to
Running. If initialization is not successful, one of the above listed conditions has not
been met.
State
Idle
Idle.
Stabilize
Waiting for input temperature and output
power to stabilize.
Running
Actively autotuning.
Complete
Successful completion.
Failed
Failed due to processing error. Usually,
this is because the process model did not
converge. Try a smaller DeltaP setting.
Abort
Aborted by the user.
Table 31: Autotune States
i
Note:
When autotuning is initiated, the algorithm will stay in the
'Stabilize' state until the output power and the input temperature are
stable. Time in this state is not part of the selected timeout. If the
system is not stable, the autotuning process will stay in the
Stabilize state indefinitely. To abort, press the
Home
key.
84