Cryo-con Model 24C
Front Panel Operation
Loop Bar Chart Display
The Loop Bar Chart is a 50-segment bar chart that shows the measured output of a
selected loop output.
The bar is composed of ten blocks with five segments. Therefore, output current can
be read to an accuracy of 2%.
Note that the bar chart does not have a loop number indicator.
Some examples are:
Loop ON, zero output:
Loop OFF:
Loop ON, 50% output:
i
Note
: The Model 24C uses an independent circuit to read
current actually flowing through the load. The heater bar graph
shows this measured current. If the unit is controlling temperature,
but the bar graph indicates zero current flow, an error condition
exists, possibly an open heater.
49