Cryo-con Model 24C
Remote Programming Guide
Relay Commands
The relay subsystem includes the two auxiliary relays in the Model 24C. Using the
RELAYS commands, these relays are independently configured to assert or clear
based on the status of any of the four sensor input channels.
Relay outputs are dry-contact and are available on the rear panel of the instrument.
The user selectable display time constant filter is applied to input channel
temperature data before relay conditions are tested. The user selectable relay
deadband is also applied.
RELays?
{1
|
2
}
Relay Status Query. The two auxiliary relays available in the Model 24C are
addressed as 0 and 1. The RELAYS command can be used to query the
status of each relay where:
--
Relay is in Auto mode and is clear.
Hi
Relay is asserted by a high temperature condition.
Lo
Relay is asserted by a low temperature condition.
ON
Relay is in manual mode and is asserted.
OFF
Relay is in manual mode and is clear.
RELays
{1 | 2}
:SOURce
{A | B | C | D}
Relay Input Source. Sets or queries the source input channel for a specified
relay.
RELays
{1 | 2}
:HIGHest <setpt>
Relay High setpoint. Sets or queries the temperature setting of the high
temperature setpoint for the specified relay. Parameter <setpt> is floating-
point numeric and is in units of the controlling input channel.
RELays
{1 | 2}
:MODe
{AUTo
|
ON
|
OFF
}
Set or query the relay mode. Modes are:
Auto
Relay is controlled by enabled high and low setpoints.
ON
Relay is in manual mode and is asserted.
OFF
Relay is in manual mode and is clear.
Control
Relay is asserted whenever the controller is in Control mode.
RELays
{1 | 2}
:LOWest <setpt>
Relay Low setpoint. Sets or queries the temperature setting of the low
temperature setpoint for a specified relay. Parameter <setpt> is floating-point
numeric and is in units of the controlling input channel.
RELays
{1 | 2}
:HIENa
{
YES
|
NO
}
Relay High Enable. Sets or queries the high temperature enable for the
specified relay.
RELays
{1 | 2}
:LOENa
{
YES
|
NO
}
Relay Low Enable. Sets or queries the low temperature enable for the
specified relay.