Cryo-con Model 24C
Instrument Calibration
Instrument Calibration
Calibration of the Model 24C controller requires the use of various voltage and
resistance standards in order to generate calibration factors for the many
measurement ranges available.
Calibration is ‘Closed-Case’. There are no internal mechanical adjustments required.
The Model 24C cannot be calibrated from the front panel.
Calibration data is stored in the instrument’s non-volatile memory and is accessed
only via the remote interfaces.
A calibration procedure document may be obtained by contacting Cryo-con technical
support at
.
Cryo-con Calibration Services
When the controller is due for calibration, contact Cryo-con for low-cost recalibration.
The Model 24C is supported on our automated calibration systems which allow Cryo-
con to provide this service at competitive prices.
Calibration Interval
The Model 24C should be calibrated at a regular interval determined by the
measurement accuracy requirements of your application.
A 90-day interval is recommended for the most demanding applications, while a 1-
year or 2-year interval may be adequate for less demanding applications. Cryo-con
does not recommend extending calibration intervals beyond 2 years.
Whatever calibration interval is selected, Cryo-con recommends that complete re-
adjustment should always be performed at the calibration interval. This will increase
user confidence that the instrument will remain within specification for the next
calibration interval. This criterion for re-adjustment provides the best measure of the
instrument’s long-term stability. Performance data measured using this method can
easily be used to extend future calibration intervals.