Cryo-con Model 24C
Specifications, Features and Functions
Sensor Selection
:
Front Panel or remote interface. There are no internal
jumpers or switches.
Sample Rate
:
15Hz per channel in all measurement modes.
Digital Resolution
:
24 bits.
Measurement Filter:
0.5, 1, 2, 4, 8, 16, 32 and 64 Seconds.
Calibration Curves
:
Built-in curves for industry standard sensors plus eight
user curves with up to 200 entries each. Interpolation is performed using a
Cubic Spline.
CalGen:
Calibration curve generator fits any diode or resistor sensor curve at
1, 2 or 3 user specified temperature points.
Sensor Performance Specifications
:
Diode Sensors
Configuration:
Constant-Current mode, 10
m
A ± 0.05% DC excitation.
Note: Current source error has negligible effect on measurement accuracy.
Input voltage range:
0 to 2.00VDC.
Accuracy:
±(80
m
V + 0.005% * reading)
Resolution:
2.3
m
V
Drift:
25ppm/ºC over an ambient temperature range of 25ºC± 5ºC.
PTC Resistor Sensors
Configuration:
Constant-Current AC resistance bridge mode.
Ratiometric measurement cancels any error in excitation current.
Drift:
20ppm/ºC over an ambient temperature range of 25ºC± 5ºC.
AC Excitation Frequency:
7.5Hz bipolar square wave.
Range
Max/Min
Resistance
Excitation
Current
Resolution
Accuracy
PTC100
1mA
500
W
0.01
W
1.0mA
0.1m
W
± (0.004 + 0.01%)Ω
PTC1K
100
m
A
7.5K
W
0.1
W
100
m
A
1.0m
W
± (0.05 + 0.02%)Ω
Table 6: Accuracy and Resolution for PTC Resistors
Note:
The Model 24C is calibrated with AC excitation. User
selection of DC excitation will introduce offset errors in temperature
measurement.
22