Cryo-con Model 24C
Introduction
Relays:
The Model 24C has two 10-Ampere dry-contact relays. These can be used
to control a refrigerator system or other external equipment.
Each relay can be asserted or cleared based on the temperature reading of a
selected input channel. High and low setpoints may be set from the front panel or a
remote interface. Furthermore, the relays can be manually asserted ON or OFF.
Remote Control
Standard Remote Interfaces
include Ethernet and RS-232. IEEE-488.2(GPIB) and
USB are optional.
The Model 24C connects directly to any Ethernet Local-Area-Network (LAN) to make
measurements easily and economically. TCP/IP and UDP data port servers brings
fast Ethernet connectivity to all common data acquisition software programs including
LabView
. An ASCII text based command language identical to those commonly
used with GPIB or RS-232 interfaces is implemented. This is the primary way that
user software interfaces to the instrument.
Using the Ethernet SMTP protocol, the controller will send e-mail based on selected
alarm conditions. E-mail is configured by using the web page interface.
Using the Ethernet HTTP protocol, the instrument’s embedded web server allows the
instrument to be viewed and configured from any web browser.