![Cryo-con 24C Скачать руководство пользователя страница 185](http://html.mh-extra.com/html/cryo-con/24c/24c_user-manual_2691809185.webp)
Cryo-con Model 24C
Appendix G: Sensor Data Tables
Thermocouples
An external thermocouple module is
required.
Thermocouple Type E
Name: TC type E
Config: TC70
K
μV
μV/K
3.2
-9834.9
1.59
4.2
-9833
2.09
10
-9813.3
4.66
20
-9747
8.51
30
-9643.8
12.1
40
-9505.5
15.5
50
-9334.2
18.7
75
-8777.7
25.6
100
-8063.4
31.4
150
-6238.1
41.2
200
-3967.4
49.3
250
-1328.7
56
273.15
0
58.5
300
1608
61.1
350
4777.7
65.6
400
8159.8
69.6
500
15426
75.3
600
23138
78.6
670
28694
80
700
31100
80.4
800
39179
81
900
47256
80.4
1000
55247
79.3
1100
63119
78.1
1200
70842
76.3
1270
76136
75.2
Thermocouple Type K
Name: TC type K
Config: TC70
K
μV
μV/K
3.2
-6457.7
0.74
4.2
-6456.9
0.92
10
-6448.5
2.01
10.5
-6447.4
2.12
20
-6417.8
4.15
30
-6365.1
6.39
40
-6290
8.61
50
-6193.3
10.7
75
-5862.9
15.6
100
-5417.6
19.9
150
-4225.5
27.5
200
-2692.8
33.5
250
-897.6
38
273.15
0
39.4
300
1075.3
40.6
350
3135.8
41.5
400
5200
40.8
500
9215.6
40.3
600
13325
41.7
670
16264
42.2
700
17533
42.4
800
21789
42.6
900
26045
42.4
1000
30251
41.7
1100
34373
40.7
1200
38396
39.7
1270
41153
39
1300
42318
38.7
1400
46131
37.5
1500
49813
36.1
1600
53343
34.5
1640
54712
34
179