Cryo-con Model 24C
Appendix C: Troubleshooting Guide
Appendix C: Troubleshooting Guide
Error Displays
Display
Condition
Or, an erratic display of
temperature.
Input channel voltage measurement is out of range.
Ensure that the sensor is connected and properly wired.
Ensure that the polarity of the sensor connections is correct. Refer
to the
Sensor Connections
section.
Many sensors can be checked with a standard Ohmmeter. For
resistor sensors, ensure that the resistance is correct by
measuring across both the Sense and Excitation contacts. For a
diode sensor, measure the forward and reverse resistance to
ensure a diode-type function.
Input channel is within range, but measurement is outside the
limits of the selected sensor’s calibration curve.
Check sensor connections as described above.
Ensure that the proper sensor has been selected. Refer to the
Input Channel Configuration Menu
section.
Change the sensor units to Volts or Ohms and ensure that the
resulting measurement is within the selected calibration curve.
Refer to the section on
Input Channel Configuration Menu
to
display the calibration curve.
151