Cryo-con Model 24C
A Quick Start Guide to the User Interface.
Configuring the Control Loops
Before using the Loop #1 (main heater) control output, it is essential that the proper
load resistance and output range be selected. This is done using the Control Loop
Setup menu as follows:
Press the
Loop 1
key.
In the Control Loop Configuration menu, Use the up, down, right and left
keys to scroll to the
Htr Resistance
field. An example is shown here:
Use the
+
and
0
keys to select between a 50-
Ohm and a 25-Ohm heater and then press
the
Enter
key.
Use the navigation keys to scroll to the Range
field and then select the desired heater range.
Be sure to select a range that does not exceed the ratings of your cryostat. A
summary of full-scale output power for the various ranges is given here:
Range
Max. Output Power
25
50
Hi
25 Watts
50 Watts
Mid
2.5 Watts
5.0 Watts
Low
0.25 Watts
0.50 Watts
Table 3: Loop #1 Output Summary
Next, the control type should be set by scrolling to the Type field and selecting the
desired loop operating mode.
+
+
18
+