Cryo-con Model 24C
Front Panel Menu Operation
The Auto Tune Menu
The Model 24C can automatically tune both control loops. For a complete description
of the autotune process including configuration of the tuning menus, refer to the
section titled
autotuning
.
The autotuning menu entries are shown below:
Auto Tune Menu
+Autotune: Loop 1
Sets the loop number for autotuning. Each control loop must be
tuned separately.
#DeltaP: 5%
Sets the maximum power delta allowed during the tuning
process. Value is a percent of full-scale output power for the
selected loop.
Mode: PI-
Sets autotuning mode. Choices are P, PI or PID.
#Timeout: 180S
Sets the autotune timeout in seconds. If the process model has
not converged within this time, tuning is aborted.
Idle Go
Pressing
Enter
will initiate the autotune sequence. The current
auto tune state is also shown.
#P=
Proportional gain term generated by autotune. This field will be
blank until a successful autotune is completed
#I=
Integral gain term generated by autotune. This field will be
blank until a successful autotune is completed.
#D=
Control Type. Selections are: Off, Man, PID, RampP and Table.
Save & Exit
Derivative gain term generated by autotune. This field will be
blank until a successful autotune is completed.
Table 26: Auto Tune Menu
69