- close the
door 16
.
- press ON and enter the access code.
- press MENU from STAND BY mode, select Menu Cycle (number 1) and press ENTER.
- use the
↑↓
arrows to select the Vacuum test cycle and press ENTER to return to STAND BY.
- press START from STAND BY:
- wait for the test to finish, shown on the display by the words “VACUUM TEST CONSISTENT”:
- press STOP, enter the personal access code and open the
door 16
.
If “REPEAT TEST” appears on the display, the test will be interrupted automatically.
Before repeating the test, open the
door 16
, let the
chamber 09
cool completely, and
check that it is perfectly dry.
Whenever you want to run the VACUUM TEST with a chamber temperature above 40°C,
the display will show a message reading “Cooling pause”, and the cycle will begin only
when the temperature drops below 40°C.
WARNING: for this reason, selecting the Vacuum test disables preheating when
the door is closed and the handle clicks.
!
If the message “VACUUM TEST NOT CONSISTENT” ever appears after running the
vacuum test, contact a TECHNICAL SERVICE CENTER.
!