GENERAL SAFETY CONDITIONS
•
Switch off the sterilizer at the
main switch 13
and disconnect the
power cable 03
before conducting any operations apart from those in the normal operations cycle.
•
The main power supply must be properly grounded.
•
It is essential that operators are properly educated. It is therefore compulsory for
them to read and comply with the technical information provided in the manual
and attached documentation. The use of unqualified personnel is strictly
prohibited.
•
Personnel who operate the sterilizer must always use the appropriate individual
protective devices: gloves for sharp or pointed objects and heated elements, and
the special tray handle for emptying the chamber.
•
Avoid any temporary repairs: repairs must be performed exclusively with original
spare parts, and these must be installed by an authorized technician.
•
Never remove or dismantle any part from the sterilizer. Only authorized personnel
trained in advance by
COMINOX
may remove the exterior protective cover and
access the interior parts.
•
Labeling plates on the unit should not be removed, damaged, or dirtied. They must
be kept clean and clearly visible.
•
Do not use the sterilizer if there are possible defects in its operation.
•
The sterilizer should be cleaned using suitable means and detergents that will not
damage or corrode any sterilizer components. It is strictly forbidden to clean or
wash the sterilizer with jets of water.
•
Never place containers of liquids on the sterilizer or the shelves above it. Leakage or
spillage on the electric system can cause a short circuit.
•
Do not place flammable or toxic materials and products for sterilization in the
chamber
09
.
•
Do not use the sterilizer to process containers that hold fluids, whether they are
sealed or not.
•
Never lean on the
door 16
.
The information in this manual does not replace any safety instructions or technical data
for installation and operation on the unit itself, nor any safety standards in force in the
installation country, or the rules of common sense.
Thermal danger
Interior parts of the
door 16
and of the
chamber 09
can reach extremely high
temperatures.
Never touch the heated surfaces until they have cooled, and always use the appropriate
protective gloves. The same applies when a load is to be removed from the sterilization
chamber. To remove the
trays 04
holding the load, always use the
tray handle 01
. Never
leave the tray handle inside the sterilization chamber.
The temperature in the sterilization chamber is controlled by a
safety thermostat
17
that
intervenes if overheating occurs. The safety thermostat must be reset manually.
Noise
The sound pressure level (
noise
) detected in suitable work spaces must be lower than
70dB(A).
General safety conditions
When opening the door 16, steam may escape, which could cause burns.
Never stand in front of or above the door!
!
Содержание SterilClave 18 VLS
Страница 2: ...2...
Страница 7: ...7 Graphic representation of references Mod 18 Table of contents...
Страница 8: ...8 Graphic representation of references Mod 24 Table of contents...
Страница 9: ...9 06 06A OPTIONAL OPTIONAL Table of contents...
Страница 10: ...10...
Страница 70: ...70...
Страница 71: ...71...