25
- get the
metering funnel 05
, insert the plug provided, and fill it based on the model with:
SterilClave 18 = 60 cc.
SterilClave 24 = 80 cc.
- use ONLY distilled or demineralized water (see Water supply table).
- slowly pour the contents of the funnel into the bottom of the chamber.
- the water cannot descend through the drainage hole, so it will flow toward the outside
of the chamber. The water must not spill out of the chamber. If the levelling is correct,
when the full quantity of water measured is poured into the chamber, it must touch the
chamber probe 21
.
- if the levelling is not correct, loosen the clamping nuts and adjust the
adjustable feet 22
.
- raise the feet if the water tends to exit. Otherwise, lower them.
- after levelling, tighten the clamping nuts on the
adjustable feet 22
, remove the water,
and dry the chamber with a cloth.
- remove the
level testing pipe 27
and replace the
waste filter 20.
- insert the
tray support 08
and the
coil covering grid 19
. Fasten with the corresponding
screw, paying attention to the
grid seal 28
.
CONNECTIONS
Electrical connections
To connect to the power supply, simply plug in the sterilizer by inserting the
power cable
03
plug into a suitable socket (see GENERAL CONDITIONS and the Technical specifications
table).
Water connection
This option comes standard on BHD and can be requested for other models.
This option makes it possible to avoid stopping the unit to fill the
clean water tank 10
and
to empty the
waste water recovery tank 11
.
1) Auto water fill
This is an alternative configuration to manually refilling the
clean water tank 10
(see
MANUALLY FILL CLEAN WATER TANK).
A)
FROM EXTERNAL CAN (maximum height differential 1 m): the
auto water fill 29
tube
can be connected directly to a can of demineralized or distilled water purchased
on the market, or to a can of distilled water produced by a Cominox water
demineralization system (Discom).
B)
FROM WATER MAIN: the
auto water fill 29
tube can be connected directly to a
Cominox water demineralization system (Speedy Water or Osmosis).
For both the
A
and
B
configurations, the Self-filling setting must be ON in the Main Settings
Menu. Configuration
B
requires installation by a qualified technician.
Check that the water quality complies with the water supply table.
WARNING: NEVER REMOVE THE FEET COMPLETELY
Do not block the space under the sterilizer.
!
Installation
Содержание SterilClave 18 VLS
Страница 2: ...2...
Страница 7: ...7 Graphic representation of references Mod 18 Table of contents...
Страница 8: ...8 Graphic representation of references Mod 24 Table of contents...
Страница 9: ...9 06 06A OPTIONAL OPTIONAL Table of contents...
Страница 10: ...10...
Страница 70: ...70...
Страница 71: ...71...